+8615966019173

fred@dmthermo.com

Low Temperature Blackbody Furnace

Home - Products - Black Body Infrared Calibration Furnace - Low Temperature Blackbody Furnace - DY-HT1D-100C Ultra-Low Temperature Blackbody Radiation Source

  • DY-HT1D-100C Ultra-Low Temperature Blackbody Radiation Source

  • DY-HT1D-100C Ultra-Low Temperature Blackbody Radiation Source

  • DY-HT1D-100C Ultra-Low Temperature Blackbody Radiation Source

  • DY-HT1D-100C Ultra-Low Temperature Blackbody Radiation Source

  • DY-HT1D-100C Ultra-Low Temperature Blackbody Radiation Source

DY-HT1D-100C Ultra-Low Temperature Blackbody Radiation Source

In daily life, we are mostly exposed to temperatures above 0°C, and ordinary blackbody furnaces (ranging from room temperature to high temperatures) are sufficient for our needs, such as calibrating industrial furnace temperature measuring instruments and clinical thermometers. However, in many cutting-edge fields, "low temperature" or even "ultra-low temperature" is the norm, requiring the reliance on cryogenic/ultra-low temperature blackbody furnaces.

For example, spacecraft encounter cryogenic environments in space, and equipment carried on board must be calibrated on the ground beforehand. Ultra-low temperature blackbody furnaces can simulate the radiation characteristics of these cryogenic environments, ensuring the reliability of measurement data after launch.


Product Introductiondry of Block Calibration Bath

Black body in the industry is mainly used in the field of temperature measurement, the main product is black body furnace.

In the calibration and verification of radiation thermometers, the comparative method is usually adopted, that is, through the radiation source of high stability (usually black body radiation source) and other supporting equipment, the temperature reproduced by the standard apparatus and the temperature reproduced by the radiation thermometer are compared, in order to judge whether it is qualified or to give the calibration results.

As a standard radiation source, the standard black body is mainly used to calibrate radiation thermometers, infrared thermometers and radiation temperature sensing detectors.

At present, our company has advanced black body technology, full product range and wide temperature range of black body series products.



Product Feature of Blackbody Radiation Source

1) Easy to use, small in size and lightweight, and portable, suitable for both laboratory and field calibration.

2) Rapid cooling and short stabilization time, suitable for multiple data tests in high and low temperature ranges.

3) Short heating and cooling time; room temperature drops to -100°C in just 40 minutes.

4) Built-in multi-point independent temperature correction function on the screen.

5) Isothermal blackbody cavity with only one cavity opening; cavity opening radiation is uniformly distributed along the spectrum, with an emissivity above 0.995 and a resolution of 0.01.

6) Automatic intelligent adjustment: Includes MH logic intelligent adjustment function and advanced control algorithms with parameter self-tuning function.

7) Programmable control function; the device has 10 different program setting interfaces, one-button operation, convenient and fast, meeting various user testing point requirements and improving user testing efficiency.

8) Communication interface can be customized; RS232 is standard.


Technical Indicators of Blackbody Radiation Calibration
ModelDY-HT1D-100C
Temperature Range-100℃-40℃
Control ModeTouch Screen
Temperature Resolution0.1℃/0.01°C
Temperature ControlPID Automatic Control
Cavity DiameterΦ40mm
Emissivity≥0.995
Temperature Stability±(0.1~0.2)@30°C
Temperature Uniformity±(0.1~0.3)@30°C
Power Supply220V AC 50/60HZ
Power800W
RemarkFor low-temperature blackbody furnaces, when the set temperature is lower than the ambient temperature, in order to effectively avoid the problem of frost formation on the radiating surface of the blackbody furnace, the blackbody furnace is equipped with a nitrogen gas supply pipeline. Users need to provide their own nitrogen, nitrogen pressure reducing valve and other related supporting equipment.
Programmable functionsThe device features programmable control with 10 different program setting interfaces. One-click operation is convenient and quick, meeting various testing point requirements and improving testing efficiency.
Dimension(L×W×H)(280×370×365)mm
Weight18Kg

Note: other special requirements can be customized according to the needs of users


Picture of Infrared Temperature Calibration

DY-HT4 Blackbody Furnace(800℃-1600℃)

DY-HT4 Blackbody Furnace(800℃-1600℃)

DY-HT4 Blackbody Furnace(800℃-1600℃)



Tags